13. A patients prescription calls for a prescribed dose to a point in the patient at a depth of 10 cm from the skin surface.

Show transcribed image text

Transcribed Image Text from this Question

13. A patient’s prescription calls for a prescribed dose to a point in the patient at a depth of 10 cm from the skin surface. This is an SAD treatment, meaning that the prescribed dose is to a point at a distance of 100 cm from the linac radiation source. A setup error is made, and the patient is actu- ally set up at 95 SSD instead of 90 SSD. Will the actual delivered dose be higher or lower, and by approximately how much? Give the answer as a percentage. 1